Rosemount™ 222 Toroidal Flow Through Conductivity Sensor
rosemount 222 conductivity sensor
The Rosemount 222 toroidal flow through conductivity sensor measures conductivity in highly conductive liquids up to 2 S/cm (2,000,000 μS/cm).
The minimum conductivity is about 500 μS/cm. A noninvasive flow through design allows for in-line installations without any obstruction to sample and makes these sensors ideal for use in viscous or fibrous liquids.
Rosemount™ 222 Toroidal Flow Through Conductivity Sensor

Specs

Ambient Temperature
150 lb Flange: 41 °F (5 °C) - 360 °F (182 °C)/125 psig (963 kPa [abs])
300 lb Flange: 41 °F (5 °C) - 360 °F (182 °C)/250 psig (1825 kPa [abs])
Dimensions
Cable Length: 20 ft (6.1 m)

Features

  • Robust measurements - insensitive to process flow rate and direction
  • Avoid accumulation of solids with no part of the sensor protruding into the sample flow
  • Corrosion resistant Teflon pipe liner
  • Versatile installation options allows for direct mounting into 1 inch and 2 inch process lines with 150 lb or 300 lb raised face threaded ANSI B16.5 flange
  • Plugging resistant and suitable for liquids containing high levels of suspended solids
  • Maximum Temperature 360°F (182°C)